8-2-51 Read and Write Serial Info
8-2-51-1 How to start serial no read/write program
1.) Plug the service dongle on the usb port on rear panel.
2.) Turn on the system.
3.) When the Start Application dialog box appears, press [F2] before it automatically starts
Echoloader.)
4.) When the Maintenance Access dialog box pops up, type in the service password.
Figure 8-152 Start Application Dialog Box
Figure 8-153 Maintenance Access Dialog Box