DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
Section 10-1 - Overview 10-1
Chapter 10
Care & Maintenance
Section 10-1
Overview
10-1-1 Periodic Maintenance Inspections 
It has been determined by engineering that your LOGIQ™ P5 system does not have any high wear 
components that fail with use, therefore no Periodic Maintenance Inspections are mandatory. Some 
Customers Quality Assurance Programs may require additional tasks and or inspections at a different 
frequency than listed in this manual.
10-1-2 Purpose of Chapter 10
This chapter describes Care & Maintenance on the scanner and peripherals. These procedures are 
intended to 
maintain the quality of the ultrasound systems performance. Read this chapter 
completely and familiarize yourself with the procedures before performing a task.
Table 10-1    Contents in Chapter 10
Section Description Page Number
10-1
Overview
10-1
10-2
Why do Maintenance
10-2
10-3
Maintenance Task Schedule
10-2
10-4
Tools Required
10-4
10-5
System Maintenance
10-7
10-6
Electrical Safety Tests
10-12
10-7
When There's Too Much Leakage Current...
10-28
Practice good ESD prevention. Wear an anti–static strap when handling electronic parts and 
even when disconnecting/connecting cables.
THERE ARE SEVERAL PLACES ON THE BACKPLANE, THE AC DISTRIBUTION, 
AND DC DISTRIBUTION THAT ARE DANGEROUS. BE SURE TO DISCONNECT 
THE SYSTEM POWER PLUG AND OPEN THE MAIN CIRCUIT BREAKER 
BEFORE YOU REMOVE ANY PARTS. BE CAUTIOUS WHENEVER POWER IS 
STILL ON AND COVERS ARE REMOVED.
Do not pull out or insert circuit boards while power is ON.
Do not operate this unit unless all board covers and frame panels are securely in place. System 
performance and cooling require this.