GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL 
8-134 Section 8-4 - Software Installation
8-4-1              Installing Base Image Software (cont’d)
8.) Press Enter key and then the following screen will be shown. (If you want to abort installation 
procedures, press CTRL+C.)
9.) Press 
A key to install the complete disk.
Figure 8-185   Base Image Software Installation Menu
Select A when installing the new HDD or upgrading from BT09. (HDD capacity should be minimum 
120GB)
Select 
B when replacing C:\ partition only. It means all customer data are NOT deleted. The Ghost pro-
cedure is performed with data of the drivers D, E, F and Z left.
Select 
C when replacing C:\ partition and Z:\ partition. It means all customer data are NOT deleted. The 
Ghost procedure is performed with data of the drivers D, E and F left. Z:\ partition used for reloading 
C:\ partition and Application software will be replaced.
Drive Partition Usage
(GRUB) grub data
C System Files and Application
D User Defined and Preset 
E Patient Archive data
FSwap
Z  C:\partition Repository for Reloading Function