GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3 - 40 Section 3-9 - Optional Peripherals/Peripheral Connection
3-9-1-2 WIth Off-board UVC
Figure 3-27 Connection Scheme - Secondary LCD Monitor with Off-board UVC
Secondary Monitor
A
ppropriate
Transformer
A Secondary “Patient” Monitor MUST NEVER be connected to the LOGIQ S8’s mains supply directly!
CAUTION
!! CAUTION:
Output connection
Following Output formats are available:
- DVI
- D-SUB (VGA)
- S-VIDEO
- COMPOSITE (BNC)
- COMPOSITE (RCA)
If the secondary patient monitor to be located inside of
isolation transformer.
the patient environment, to be connected to the appropriate