EasyManua.ls Logo

GE LOGIQ S8 R3

GE LOGIQ S8 R3
418 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips 7 - 71
7-7-5 Trackball - Impaired sensitivity
Figure 7-64 Trackball - Impaired sensitivity
Trackball has low sensitivity.
Set up the appropriate trackball response time.
(via [Speed] and [Acceleration] in the Utility - System - General
- Trackball. Refer to Figure 3-23 on page 3-33)
Is the trackball issue resolved?
Is the trackball issue resolved?
Clean the trackball.
Refer to section 7-7-5-1 on page 7-72
Replace the trackball.
Return to normal scanning. Return to normal scanning.
No
No
No
Yes
Yes

Table of Contents

Related product manuals