GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
3 - 78 Section 3-12 - Configuring Connectivity
3-12-7-2 Removing a service
1.) Select the service. Press Remove.
2.) Press Save.
3-12-7-3 Changing parameters for a service
There are certain parameters that may need to be set up for each service:
Many service parameters are specific to each type of service. The parameters are described on the
following pages:
DICOM Image Storage
DICOM Performed Procedure
DICOM Print
DICOM Query/Retrieve
DICOM Storage Commitment
DICOM Worklist
Standard Print
Video Capture
Save As
Network storage
USB Quick Save (R2 and later)
Email to MMS (R4 and later)
Table 3-12 Service Parameters: Common Service Parameters
Preset Parameter Description
Name Free text: give a descriptive name to the device.
AE Title The Application Entity Title for the service.
Port Number The port number of the service.
Maximum Retries Max # - the maximum number of times to try establishing a connection to the service.
Retry Interval (sec.) Specify how often (in seconds) the system should try to establish a connection to the
service.
Timeout The amount of time after which the system will stop trying to establish a connection
to the service.