DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-3 - Gathering Troubleshooting Data (R1 to R4) 7 - 19
7-3-7 Capturing Network Logs with Network Sniffer (cont’d)
14.) Select Displayed. This will save only the filtered values rather than the entire capture.
15.) Select SAVE.
If you perform Alt+D or Gather Logs, these sniffer logs will be included in the zip file.
Figure 7-19 Select Capture Display
16.)Exit the sniffer application by clicking on the X in the upper right corner.
17.)If you have already performed Save as, you can continue without saving and quit the program.
Make sure you have saved your data before exiting the program.
Figure 7-20 Save Capture Data Question