GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 7-6 - Service Desktop (R4.2.5x and later) 7 - 55
Change Password
Change Password allows you to change the password for a specified user type.
The information on Change Password is available to all service class licenses. Change Password is
not available through a remote connection.
To access Change Password, select Utility (second page) > Service > Utilities > Change
Password.
Figure 7-53 Change Password
This table shows all the elements available on Change Password with descriptions.
CAUTION
RISK OF LOSING DATA.
IF THE PASSWORD IS LOST, GE WILL NOT BE ABLE TO RECOVER OR RESET IT.
LOSS OF A PASSWORD MAY RESULT IN THE LOSS OF PATIENT DATA.
Table 7-7 Change Password
Element Description
User Type
Type of user for the password reset.
• GEService – Windows user for GE Service.
• SVCService – Windows user loaded with the base image load software.
New Password
New password according to the password requirements (14-25 characters including
one upper cap, one lower cap, one number, and one symbol).
Confirm Password Password you entered in New Password.
Update Password Select to update the password.
Reset Select to reset the information.