GE HEALTHCARE
DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
7 - 80 Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips
7-7-16-1 Note for FIbroscan
1.) Sleep mode
FibroScan may not start after sleep mode because the time for booting the FibroScan is not enough.
Do not use sleep mode when using FibroScan.
2.) Pointint cursor
Pointing cursor is available for only the following red area of the screen during FibroScan Mode.
3.) Measurements in Report/Worksheet
Order of the measurements isn't equal between FibroScan Report and LS8 Report/Worksheet.
However, measurements results are correct.
7-7-16-2 Issues
1.) CAP and KPa values occasionally not displayed.
At the 1st time of FibroScan mode after system starts, CAP score and Fibrosis KPa values
occasionally not displayed.
<RECOVERY>
Once exit from FibroScan mode, and enter to FibroScan mode again.