DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
7 - 92 Section 7-8 - Error messages
Information “Starting Battery Operation”
Condition:
After unplugged and waiting to enter battery power mode (3 seconds).
Information “Running on Battery. Key operation locked.”
Condition:
Battery power mode (power saving, locked keys, etc.)
No message Condition:
Plugged.
Warning “Battery has deteriorated. Replace the battery for proper function.”
Condition:
Battery has deteriorated. Shown periodically (14 days) on startup.
Warning “Low Battery. System shutting down.”
Condition:
Shutdown due to low battery power.
Warning “Error in Battery. System shutting down.”
Condition:
Shutdown due to battery error.
Warning “Battery cannot be detected.”
Condition:
Shutdown due to battery error.
Table 7-20: Error messages