DIRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 9-8 - Electrical Safety Tests 9 - 23
9-8-6 Data sheet for enclosure/chassis leakage current
Table 9-15 below shows a typical format for recording the enclosure/chassis leakage current.
Measurements should be recorded from multiple locations for each set of test conditions.
The actual location of the test probe may vary by Ultrasound system.
NOTE: Record all data in the Electrical safety tests log.
NOTE: Values in italics font are given as examples only.
Table 9-15 Typical data format for recording enclosure/chassis leakage
Unit under test___________________________ Date of test:_____________________
Test Conditions Measurement/Test Point Location
System
Power Grounding/PE
Rear
Panel
Lower
Frame
Probe
Connector
Main
Handle
Off Closed
Off Open
On Closed
On Open