Maintenance
ElectricalSafetyChecks(cont'd.)
Step
PolarityCondition
1
UUT
2
-ON
ResultLeakage
Current
Limits
1
Forward
Polarity
NC
______µA
Pass/Fail
100µA
2
Neutralopen,
Forward
Polarity
SFC
______µA
Pass/Fail
300µA
3
Groundopen,
Forward
Polarity
SFC
______µA
Pass/Fail
300µA
4
Groundopen,
Reverse
Polarity
SFC
______µA
Pass/Fail
300µA
5
Neutralopen,
Reverse
Polarity
SFC
______µA
Pass/Fail
300µA
6
Reverse
Polarity
NC
______µA
Pass/Fail
100µA
PatientLeakageCurrenttoGround
1
Forward
Polarity
NC
______µA
Pass/Fail
10µA
2
Neutralopen,
Forward
Polarity
SFC
______µA
Pass/Fail
50µA
3
Groundopen,
Forward
Polarity
SFC
______µA
Pass/Fail
50µA
4
Groundopen,
Reverse
Polarity
SFC
______µA
Pass/Fail
50µA
5
Neutralopen,
Reverse
Polarity
SFC
______µA
Pass/Fail
50µA
6
Reverse
Polarity
NC
______µA
Pass/Fail
10µA
GroundContinuity
Resistance
2028451-183EMAC™1600103