Maintenance
DiagnosticTests
1.Verifythatthedisplaytestwassuccessful.
Referto“TestingtheDisplay”onpage35formoreinformation.
2.Verifythatthekeyboardtestwassuccessful.
Referto“TestingtheKeyboard”onpage37formoreinformation.
3.Verifythatthewritertestwassuccessful.
Referto“TestingtheWriter”onpage39formoreinformation.
4.Verifythattheacquisitionmoduletestwassuccessful.
Referto“TestingtheAcquisitionModule”onpage38formoreinformation.
5.Verifythatthepatientleadwirecheckwassuccessful.
Referto“TestingthePatientLeadWires”onpage46formoreinformation.
6.Verifythatthebatterystatusisacceptable.
Referto“CheckingBatteryStatus”onpage38formoreinformation.
7.VerifythattheLANtestwassuccessful.
Referto“TestingtheLANOption”onpage43formoreinformation.
8.Verifythatthemodemtestwassuccessful.
Referto“”onpageformoreinformation.
ElectricalSafetyChecks
Verifythatthecurrentleakagetestresultsmeetrequirements.
Performelectricalsafetycheckswhenindicated.Allindicatedelectricalsafetychecks
requireapass/failindicationforthestepsperformed.Recordthemeasurement
valuesinyourdebrief.
ElectricalSafetyChecks
Step
PolarityCondition
1
UUT
2
-ON
ResultLeakage
Current
Limits
EarthLeakageCurrent
1
Forward
Polarity
NC
______µA
Pass/Fail
300µA
2
Neutralopen,
Forward
Polarity
SFC
______µA
Pass/Fail
1,000µA
3
Neutralopen,
Reverse
Polarity
SFC
______µA
Pass/Fail
1,000µA
4Reverse
Polarity
NC
______µA
Pass/Fail
300µA
EnclosureLeakageCurrent