PRODUCT OR
COMPONENT
NET WEIGHT
KG (LBS)
DIMENSIONS MM (INCHES) LOAD BEAR‐
ING AREA
MM (IN‐
CHES)
WEIGHT/ OC‐
CUPIED
AREA
MOUNTING METHOD
WIDTH DEPTH HEIGHT
DL LCD monitor 8.2 (18) 179 (7) 387 (15.2) 504 (19.8)
VCIM 0.95 (2.09) 450 (17.7) 150 (5.9) 50 (2)
Pre-cabled Ergo
LCD monitor sus‐
pension (self
weight without
monitor and ac‐
cessories given)
94 (208)
See Illustration
Ergo LCD suspension
dimension
in Dimension Drawings
Ceiling-mounted
1 pair of Stationary
rails (include cable
track) for Ergo
Suspension
33 (73)
See Illustration
Ergo LCD suspension
dimension
in Dimension Drawings
Ceiling-mounted
NOTE: (1) including patient weight (patient weight considered is 204 Kgs/450 lbs).
NOTE: (2) Depth.
NOTE: (3) Maximum dimensions given. Exact dimensions depend on chiller manufacturer
type.
NOTE: (4) Recommended supplier.
3.3.2 Positioner and Table Floor Mounting
The distances between the LC Positioner and the Tables are critical for a proper clinical usage.
For this reason, GEMS provides two floor mounting templates to ensure these components are
properly placed in relation to one another.
Table 2-8:
Title Illustration
LC Positioner Floor Mounting Methods Illustration 2-58and Illustration 2-59
Cable Conduit For On-Grade Floor Anchor Kit Illustration 2-60
Inner Base Plate For Above Grade Floor Anchor Kit Illustration 2-61
LC Positioner Floor Mounting Template
Illustration 2-63 and Holes location in concrete floor il‐
lustration in Floor Requirements.
Fixing Bolt Overview Illustration 2-62
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
114 3 Room Structural Requirements