4 Mounting Data, Including Seismic
4.1 Seismic areas
4.1.1 Floors
Chiller: The seismic kit for the Chiller 4100 (SMC) is included with the chiller when shipped.
In Seismic areas all cabinets must be anchored to the floor. See Basic System Compatibility for
referential documents. System Control Cabinets: Wall support 5404587.
Every sub–system is delivered on site with its proper seismic kit.
•
Monitor Flat Panel Seismic Kit: 2353317
•
VCIM seismic kit: 2365510.
Anti–seismic means be installed before opening the system for normal use.
4.1.2 Walls
NOTE: For Fluoro UPS option (US version only), the Fluoro UPS supplier will deliver
anchoring brackets (see Illustration 2-68). The bolts will be provided by the customer.
Consider local seismic codes when planning cabinet mounting. Consult seismic expert to
determine which mounting method is appropriate for the seismic region. Certain seismic regions
require additional reinforcement in walls. See Basic System Compatability for referential
documents.
Illustration 2-68:
4.2 Seismic Calculations
Seismic requirements are determined and specified by the hospital/ Design Professional of
record and may require approval by the specific state or country agency.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
136 4 Mounting Data, Including Seismic