2 Altitude and Atmospheric Pressure
Refer to Table 4-2. To obtain altitude and atmospheric pressure requirements for components
not specified in Table 4-2, refer to the appropriate component Pre–Installation Manual listed in
Chapter 2, Basic System Compatibility.
Table 4-2: Altitude and Atmospheric Pressure
INSTALLATION ROOM OF PRODUCT OR
COMPONENT
ALTITUDE (meters) ATMOSPHERIC PRESSURE (kPa)
IN-USE IN-USE
MIN MAX MIN MAX
Examination room 0 2000 79.4 106
Technical room (See Notice below 0 2000 79.4 106
Control room 0 2000 N/A N/A
NOTICE
The chiller is able to dissipate maximum continuously power at 3000 meters in an
ambient temperature up to 20°C, for this altitude the technical room temperature
shall not exceed 20°C.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 4 Environmental Requirements 157