1.2.3 User Interfaces
Illustration 2-3: User interfaces with Omega table
1.2.4 Description of the X–Ray generator
The System uses a High Voltage System with component parts as follows according to the
power requested (100 kW):
Jedi – System Control Cabinet,
Ingrid H.V. tank mounted on the Performix 160A X–Ray tube housing.
1.2.5 Description of the X–Ray head
The Optima System uses a Performix 160A X–Ray tube housing:
X–Ray tube
collimator
Ingrid HV tank
oil/water exchanger
contour filter
tube chiller
NOTE: An external recirculating chiller is mandatory.
It is mandatory to place the Chiller in the Technical Room.