2.2 MIS (Master Interconnect System)
Optima system interconnect cables are described in MIS (Master Interconnect System)
documents. These documents specify all interconnections between components within the
system.
Reference: For specific Vascular system interconnect maps and connection details, refer to the
following
Optima IGS 320, Optima IGS 330 MIS Map
Optima IGS 320, Optima IGS 330 MIS Charts
General Guidelines
Optima System introduce a new system interconnect with a star distribution for all cables from
the technical area. The cables are shipped on spools to create cable groups. Cable group 1 for
Exam room and cable group 2 for control room. The cable group shall be put in place during the
same action. The cables are routed in the same duct.
The HV cables could be pulled separately.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 5 Electrical Requirements 185