The relative humidity shall be at least 30 percent.
○
The dissipative material shall be connected to the system ground
reference, if applicable.
NOTICE
Route separately EMI filter incoming power lines and outgoing power lines (refer
to
Electrical Connections
).
NOTE: The maximum distance between the EMI filter and the PDB is 3 m.
NOTE: The Fluoro UPS (CE) option requires an EMI filter box placed upstream the PDB. The
EMI filter box will be delivered as a single product by GE Consumer & Industrial.
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
154 2 EMI Consideration