EasyManua.ls Logo

GE Optima IGS 330 - Page 189

GE Optima IGS 330
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NOTE: 18 meters (59 ft) is the only cable length available for the Remote TSUI box data
cable connecting remote TSUIs in the control booth and the patient table.
NOTE: Never connect the Physio module to the ground.
The Physio cable can run in the same conduit as the Bolus cable. In this case, it is
required to have a conduit between the table and the physio gases box.
If no conduit available between rear of table and Control Room (no Remote TSUI, no
MacLab…), need to define proper cable routing or create new conduit as per PIM
requirements.
If there is no physio gases box behind the table in the lay out, find a local solution to
hide the hole in the floor and the cable exit.
On some sites there is only one conduit between Exam room and Control room for the
MacLab
Optima IGS 320, Optima IGS 330 Pre-Installation Manual
Direction 5537562-1-1EN, Revision 3
Chapter 5 Electrical Requirements 189

Table of Contents

Related product manuals