PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
11-3
TABLE 11-1
PLANNED MAINTENANCE
MAINTENANCE ITEM FREQUENCY (MONTHS) PROCEDURE
GENERAL
Visual inspection 12 months Check for debris that would indicate abnormal
wear.
General cleaning and painting As required See section 11-2-1
Functional checks 12 months Perform tasks described in Proteus XR/a
Service Manual.
TABLE BASE
Telescopic covers 12 months Check conditions and cleans.
Control pedals and movement
inhibition buttons
12 months Check condition and functioning. Ensure pedals
and buttons have free play without any obstruction.
SID display meter 12 months Check the SID display shows the real SID.
Electrical Cables 12 months Check conditions of cable sheathing and correct
routing.
Power and ground connections. 12 months Check condition and connections.
Connector panel and cable
connectors.
12 months Check condition and connections.
TABLE TOP
Surface finish. 12 months Check conditions and cleans.
Longitudinal bearings and
wheel bearings.
12 months Check condition. Lubricate with molylithium grease.
Longitudinal bearing tracks and
rails.
12 months Clean and wipe with light machine oil or WD40.
Transversal rods and bearings. 12 months Clean condition, Clean and wipe with light machine
oil. Lubricate with molylithium grease.
Rubber bumper end stops
(Longitudinal & transversal)
12 months Check condition of stops and tightness of
anchoring.
CASSETTE TRAY AND
CARRIAGE
Grid oscillating mechanism. 12 months Check that movement is not obstructed.
Electrical connector and cable. 12 months Ensure connector is in good condition and making
good contact. Check condition of cable sheathing.
Cassette handle assembly,
push button and hand holder.
12 months Ensure handle and push button has free play
without any obstruction.
Bearings and tracks (front and
rear)
12 months Check condition. Clean and wipe with light
machine oil. Lubricate with molylithium grease.
Rubber bumper end stops. 12 months Check condition of stops and tightness of
anchoring.
Tray 12 months Check general condition of the tray.
Cassette size marker and
locking assembly.
12 months Check condition. Clean and wipe with light
machine oil.
Electrical connector and cable. 12 months Ensure connector is in good condition and making
good contact. Check condition of cable sheathing.
FOR TRAINING PURPOSES ONLY!