PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
5-5
Console Message: Error 80 Hardware error
Recommended Operator Action: The Proteus XR/a system has
detected a hardware error. Press
the reset button and try the
exposure again. If error occurs
again note the error and call
service.
Console Message: Error 90 Software error
Recommended Operator Action: The Proteus XR/a system has
detected a software error. Press the
reset button and try the exposure
again. If error occurs again note the
error and call service.
Console Message: Error 100 System Communication error
Recommended Operator Action: The Proteus XR/a system has
detected a system communi-cation
error. Press the reset button and try
the exposure again. If error occurs
again note the error and call
service.
Console Message: Error 110 Tube/generator overheat error
Recommended Operator Action: The Proteus XR/a system has
detected a tube/generator overheat
error. Press the reset button and
wait until tube cooling down then try
the exposure again. If error occurs
again note the error and call
service.
Console Message: Error 120 Application error
Recommended Operator Action: The Proteus XR/a system has
detected a application error. This
maybe due to a technique selection
error or when the exposure switch
was released before the exposure
was completed. Press the reset
button and change the technique or
make sure to hold down the
exposure switch untill the exopsure
is completed, then try the exposure
again. If error occurs again note the
error and call service.