PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
5-4
5-1-3 Group 3 Error Message Screen
The Error message area of the display screen displays informational
messages to inform the user of system and subsystem operational
status. In this situation, all buttons will be inactive except the “OK”
button. The following messages will appear in the error message area.
Console Message: X-ray Room Door Open
Recommended Operator Action The door to the x-ray room is not
closed. The system will pro-hibit an
x-ray exposure until the door is
closed.
Console Message: Receptor Selection error
Recommended Operator Action This error will occur when the
selected receptor is not configured
on the Jedi generator configuration
menu.
Console Message: Error 30 Tube Spit error
Recommended Operator Action: The Proteus XR/a system has
detected a tube spit error. Press the
reset button and try the exposure
again. If error occurs again note the
error and call service.
Console Message: Error 40 Rotation error
Recommended Operator Action: The Proteus XR/a system has
detected a rotation error. Press the
reset button and try the exposure
again. If error occurs again note the
error and call service.
Console Message: Error 50 Heat (filament) error
Recommended Operator Action: The Proteus XR/a system has
detected a heat (filament) error.
Press the reset button and try the
exposure again. If error occurs
again note the error and call
service.
Console Message: Error 60 Exposure error
Recommended Operator Action: The Proteus XR/a system has
detected a exposure error. Press
the reset button and try the
exposure again. If error occurs
again note the error and call
service.
Console Message: Error 70 Power Supply error
Recommended Operator Action: The Proteus XR/a system has
detected a power supply error.
Press the reset button and try the
exposure again. If error occurs
again note the error and call
service.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).