ELECTROMAGNETIC COMPATIBILITY (EMC) (CONT.)
Note: To comply with the regulations applicable to an electromagnetic interface for a
Group 1, Class A Medical Device, and to minimize interference risks, the
following requirements shall apply:
• All interconnect cables to peripheral devices must be shielded and
properly grounded. Use of cables not properly shielded and grounded
may result in the equipment causing radio frequency interference in
violation of the European Union Medical Device directive and FCC
regulations.
• All of those recommended guidance regarding electromagnetic
environment should be followed.
Note: Do not use devices that intentionally transmit RF signals (Cellular Phones,
Transceivers, or Radio Controlled Products) in the vicinity of this equipment as it may
cause performance outside the published specifications. Keep the power to these
type devices turned off when near the equipment.
The medical staff in charge of this equipment is required to instruct technicians,
patients, and others.
Guidance and manufacturer’s declaration – Electromagnetic Emissions
The Proteus XR/a system is suitable for use in the specified electromagnetic environment. The
purchaser or user of the Proteus XR/a system should assure that it is used in an
electromagnetic environment as described below:
Emissions Test Compliance Electromagnetic Environment
RF Emissions
CISPR11
Group1
The Proteus XR/a system uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
RF Emissions
CISPR11
Class A
Harmonic emissions
IEC 61000-3-2
Not
applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not
applicable
The Proteus XR/a system is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
FOR TRAINING PURPOSES ONLY!