Revision B SEER Light Ambulatory Recorder/Controller 2-15
2019818-008
SEER Light/SEER Light Extend Recorder: Troubleshooting
Cannot Record for 48 Hours
Cannot record for
48 hours.
Are there many
artifacts in the ECG
recording?
Refer to flow chart on
artifact later in this chapter.
Were the batteries
removed before recording
for 48 hours?
Don’t remove the batteries
before recording for 48 hours.
Were alkaline
batteries used?
Use alkaline
batteries.
Are the
battery electrodes dirty
or improperly
soldered?
Clean the battery electrodes
or improve soldering.
Call GE Service.
Yes
No
No
No
Yes
Yes
No
Yes
Is the recorder
labeled “SEER Light
Extend?”
The recorder does not have
48-hour recording
capability.
No
Yes