EasyManuals Logo

GE SEER Light Extend Ambulatory Recorder User Manual

GE SEER Light Extend Ambulatory Recorder
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
3-6 SEER Light Ambulatory Recorder/Controller Revision B
2019818-008
SEER Light/SEER Light Extend Controller: Troubleshooting
No Operation After Inserting Batteries and Pressing Power Button
No operation after inserting
batteries and pressing the
POWER button.
Are the batteries
set properly?
Set the batteries
properly.
Do the
batteries have sufficient
power of >4.5 V?
Replace with new
batteries.
Did the
beep sound when the
POWER button was
pressed?
Call GE service.
No
Yes
Yes
No
No
Yes
Call GE service.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE SEER Light Extend Ambulatory Recorder and is the answer not in the manual?

GE SEER Light Extend Ambulatory Recorder Specifications

General IconGeneral
BrandGE Healthcare
ModelSEER Light Extend
ApplicationHolter Monitoring
Impedance CheckYes
Event ButtonYes
Wireless CommunicationBluetooth
Battery LifeUp to 7 days
ConnectivityUSB
Recorder TypeAmbulatory
Power SourceBattery operated
Memory CapacityUp to 7 days
CompatibilityCompatible with GE Healthcare systems
Number of Channels3 channels