Revision B SEER Light Ambulatory Recorder/Controller 3-13
2019818-008
SEER Light/SEER Light Extend Controller: Troubleshooting
Recording Start Date/Time is Not Correct
Time is Not Correct
Recording start date/time
is not correct.
Is date within
one month?
Adjust date.
See next flowchart.
Yes
No
Time is not correct.
Is it within 5 years
of use?
Call GE service.
Are the
battery electrodes
clean or properly
solder?
Clean the electrodes
or improve soldering.
Replace the PC boards.
Yes
Yes
No
No