2-18 SEER Light Ambulatory Recorder/Controller Revision B
2019818-008
SEER Light/SEER Light Extend Recorder: Troubleshooting
Record Start Date/Time is Initialized
Record start date/time
is initialized.
Was the recording
started by pressing the
START button?
Call GE Service.
Were the batteries
removed during or after
recording?
Record start date/time will be
initialized if the batteries are
removed before transferring the
ECG data after the START button
is pressed to start recording.
Replace the batteries.
Yes
No
No
Yes
Did more than
one week pass from the
last recording?
Call GE Service.
No
Yes