Revision B SEER Light Ambulatory Recorder/Controller 2-5
2019818-008
SEER Light/SEER Light Extend Recorder: Troubleshooting
Troubleshooting
Self-Test Mode
A self-test can be done on the following functions.
1. Self-test I (pacemaker detection check mode)
Provide a pacemaker signal to the recorder with an ECG simulator.
To enter the pacemaker detection mode, set the batteries while
pressing the STOP button.
Confirm the audible sound synchronizes with the pacemaker
signal input through the patient input connector.
Remove the batteries to end the self-test.
2. Self-test II (accelerometer sensor check mode)
After setting the batteries, press the STOP button three times
within a second.
Confirm the audible sound synchronizes as the recorder is moved
around.
Remove the batteries to end the self-test.