2-10 SEER Light Ambulatory Recorder/Controller Revision B 
2019818-008
SEER Light/SEER Light Extend Recorder: Troubleshooting
ECG Cannot Be Previewed with SEER Light Connect Device
ECG cannot be previewed 
with the SEER Light 
Connect Device.
Is the distance from 
the controller within 
4.9 ft. (1.5 m)?
Position the controller 
within 4.9 ft (1.5 m).
Is the I.R.
window facing the 
controller properly?
Reposition the I.R. window 
properly, or remove any 
obstacle between the 
recorder and the controller.
Were the batteries 
installed within one 
hour?
Reinstall the batteries.
Has recording 
already started?
Call GE service.
Cannot preview ECG by I.R. 
operation if more than 3 
minutes has elapsed since 
recording started.
Yes
No
Yes
Yes
No
No
No
Yes
Is the USB cable 
from the Connect 
plugged in?
Plug it in.
Yes
No
Is the orange LED 
on the Connect on?
Call GE Service.
Yes
No