Revision B SEER Light Ambulatory Recorder/Controller 3-5
2019818-008
SEER Light/SEER Light Extend Controller: Troubleshooting
Troubleshooting
Software Version
The software version of the SEER Light/SEER Light Extend Controller
appears in the lower right hand corner of the display upon start up.
Self-Test
A self-test can be done on the following functions.
1. Button check mode:
Enter the Set-up Condition display mode by pressing the POWER
button and button simultaneously. Continue to hold the button
while releasing the POWER button.
Press the CLEAR button at the Set-up Condition display mode.
Press the number 0 at the Factory Settings display mode.
The name of the button that was pressed will be displayed, and
an audible sound will be heard when the respective button is
pressed once.
NOTE
If the ENTER button is pressed three times the unit will
enter into the LCD check mode.
Turn off the power to end this function test.
2. LCD check mode:
Enter the Set-up Condition display mode by pressing the POWER
button and button simultaneously. Continue to hold the button
while releasing the POWER button.
Press the CLEAR button at the Set-up Condition display mode.
Press the number 0 at the Factory Settings display mode.
In the button check mode display press the ENTER button three
times, and the LCD display will appear darkened.
Press the POWER button to end this function test.