B-6 SEER Light Ambulatory Recorder/Controller Revision B 
2019818-008
Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Recommended Separation Distances
The table below provides the recommended separation distances (in 
meters) between portable and mobile RF communication equipment and 
the SEER Light recorder and controller.
The SEER Light recorder and controller are intended for use in the 
electromagnetic environment on which radiated RF disturbances are 
controlled. The customer or the user of the SEER Light recorder and 
controller are can help prevent electromagnetic interference by 
maintaining a minimum distance between portable and mobile RF 
communications equipment (transmitters) and the SEER Light recorder 
and controller as recommended below, according to the maximum output 
power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the 
recommended separation distance [d] in meters (m) can estimated using 
the equation applicable to the frequency of the transmitter, where P is 
the maximum output power rating of the transmitter in watts (w) 
according to the transmitter manufacturer.
NOTE
These guidelines may not apply in all situations. Electromagnetic 
propagation is affected by absorption and reflection from structures, 
objects, and people.
Separation Distance in Meters (m) According to Frequency of Transmitter
Rated Maximum Output 
Power of Transmitter in 
Watts
150 kHz to 80 MHz 
outside ISM bands
d = 1.2
150 kHz to 80 MHz
in ISM bands
d = 1.2
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5 GHz
d = 2.3
0.01 0.12 0.12 0.12 0.23
0.1 0.38 0.38 0.38 0.73
1 1.2 1.2 1.2 2.3
10 3.8 3.8 3.8 7.3
100 12 12 12 23
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.