Revision A SEER Light Ambulatory Recorder/Controller 4-5
2019818-008
SEER Light Connect: Troubleshooting
ECG Cannot Be Previewed with SEER Light Connect Device
ECG cannot be previewed
with the SEER Light
Connect Device.
Is the distance from
the controller within
4.9 ft. (1.5 m)?
Position the controller
within 4.9 ft (1.5 m).
Is the I.R.
window facing the
controller properly?
Reposition the I.R. window
properly, or remove any
obstacle between the
recorder and the controller.
Were the batteries
installed within one
hour?
Reinstall the batteries.
Has recording
already started?
Call GE service.
Cannot preview ECG by I.R.
operation if more than 3
minutes has elapsed since
recording started.
Yes
No
Yes
Yes
No
No
No
Yes
Is the USB cable
from the Connect
plugged in?
Plug it in.
Yes
No
Is the orange LED
on the Connect on?
Call GE Service.
Yes
No