4-4 SEER Light Ambulatory Recorder/Controller Revision A 
2019818-008
SEER Light Connect: Troubleshooting
Troubleshooting 
WARNING
LEAKAGE CURRENT—Electrical shock to patient could 
result from component failure and lack of power 
isolation. 
In the event this system is used in the patient vicinity, it 
must be configured in such a way that it and all of its 
electrically-connected peripheral devices are isolated from 
mains power to prevent excessive leakage current to the 
patient. This can be accomplished through the use of 
isolated mains power, or a medical grade isolation 
transformer (in compliance with UL 60601, CAN/CSA 
C22.2 No. 601.1, IEC 60601-1) with this system. All non-
medical peripheral devices shall comply with IEC and ISO 
safety standards that are relevant to that equipment (i.e., 
IEC 60950, UL 60950).
Use of the SEER Light Connect device in the patient 
vicinity requires that these measures are observed.
Software Version 
To view the software version of the SEER Light Connect device, click 
About in the SEER Light Hookup window.