B-4 SEER Light Ambulatory Recorder/Controller Revision B 
2019818-008
Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Guidance and Manufacturer’s Declaration – Electromagnetic 
Immunity
The SEER Light recorder and controller are intended for use in the 
electromagnetic environment specified below. It is the responsibility of 
the customer or user to ensure that the SEER Light recorder and 
controller are used in such an environment.
NOTE
Ut is the AC mains voltage prior to application of the test level.
Immunity Test EN 60601 Test Level Compliance Level
Electromagnetic Environment - 
Guidance
Electrostatic 
discharge (ESD) 
EN 61000-4-2
± 6 kV contact 
± 8 kV air
± 6 kV contact 
± 8 kV air
Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30%.
Electrical fast 
transient/burst 
EN 61000-4-4
± 2 kV for power supply lines
±1 kV for input/output lines
± 2 kV for power supply lines
±1 kV for input/output lines
Not applicable. The SEER Light recorder 
and controller are not powered by Mains 
power.
Surge 
EN 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Not applicable. The SEER Light recorder 
and controller are not powered by Mains 
power.
Voltage dips, short 
interruptions and 
voltage variations on 
power supply input 
lines
EN 61000-4-11
<5% U
t
 (>95% dip in U
t
)
for 0.5 cycles
40% U
t
 (60% dip in U
t
)
for 5 cycles
70% U
t
 (30% dip in U
t
)
for 25 cycles
<5% U
t
 (>95% dip in U
t
)
for 5 sec
<5% U
t
 (>95% dip in U
t
)
for 0.5 cycles
40% U
t
 (60% dip in U
t
)
for 5 cycles
70% U
t
 (30% dip in U
t
)
for 25 cycles
<5% U
t
 (>95% dip in U
t
)
for 5 sec
Not applicable. The SEER Light recorder 
and controller are not powered by Mains 
power.
Power frequency 
(50/60 Hz) magnetic 
field
EN 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be 
at levels characteristics of a typical location 
in a typical commercial or hospital 
environment.