Revision B SEER Light Ambulatory Recorder/Controller 5-3
2019818-008
Maintenance: Maintenance
Maintenance
Visual Inspection
Perform a visual inspection daily. If you notice any items that need
repair, contact an authorized GE Medical Systems Information
Technologies service person to make the repairs.
Check the case and display screen for cracks or other damage.
Regularly inspect all cords and cables for fraying or other damage.
Inspect all plugs, cables, and connectors for bent prongs or pins.
Verify that all cords and connectors are securely seated.
Inspect controls for proper operation.
Inspect the LCD to make sure all the segments function.
Precautions
Do not immerse any part of the equipment in water.
Do not use organic solvents, ammonia based solutions, or abrasive
cleaning agents which may damage equipment surfaces.
NOTE
Remove the batteries before cleaning
Cleaning
Clean the exterior surfaces with a clean, soft cloth and a mild
dishwashing detergent diluted in water.
Wring the excess water from the cloth. Avoid contact with open
vents, plugs, or connectors.
Do not spray or spill fluid directly on the recorder/controller.
Dry the surfaces with a clean cloth or paper towel.
Cleaning Frequency
With each new patient use, clean the SEER Light Recorder/Controller
and the SEER Light recorder pouch, leadwires, and patient cable.