2-12 SEER Light Ambulatory Recorder/Controller Revision B 
2019818-008
SEER Light/SEER Light Extend Recorder: Troubleshooting
Recording Cannot Start with the SEER Light Connect Device
Recording cannot start 
with the Connect.
Does the controller 
display “untransmitted 
data?”
Transfer previous data or 
press the START button.
Does
the controller display 
“Already started?”
Reinstall the batteries over 30 
seconds after removing them.
Is the distance 
from the Connect 
within 4.9 ft 
(1.5 m?)
Adjust the position 
within 4.9 ft. (1.5 m) 
from the Connect.
Does the I.R. window 
face the Connect 
properly?
Reposition the I.R. windows 
to face each other properly. 
Remove any obstacles 
between the recorder and 
the Connect.
Call GE Service.
Yes
Yes
No
No
Yes
Yes
No
No
Is the USB cable 
from the Connect 
plugged in?
Plug it in.
Yes
No
Is the orange LED 
on the Connect on?
Call GE Service.
No
Yes