General procedures
Venue 50 – Basic Service Manual 4-3
5447566-100 English Rev. 7
General procedures
Overview
Some procedures are used more often than other. The intention
with this section is to keep the most used procedures in one
place.
WARNING
Energy Control and Power Lockout for Venue 50.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
6. Remove/disconnect the battery, if present.
Ultrasound System components may be energized.