Electrical safety tests
Venue 50 – Basic Service Manual 10-25
5447566-100 English Rev. 7
Grounding continuity
Measure the resistance from the third pin of the attachment plug
to the exposed metal parts of the case. The ground wire
resistance should be less than 0.1 ohms. Reference the
procedure in the IEC60601-1.
Figure 10-3. Ground continuity test
DANGER ELECTRIC SHOCK HAZARD. THE PATIENT MUST NOT BE
CONTACTED TO THE EQUIPMENT DURING THIS TEST.
1. GROUND PIN
2. OHMMETER
3. DOCKING CART
4. ACCESSIBLE METAL PART:
• I/O Board
• Probe cable hook