System maintenance
Venue 50 – Basic Service Manual 10-17
5447566-100 English Rev. 7
Battery Performance Maintenance
Battery replacement every three years is recommended.
It is recommended to do battery performance maintenance one
time per year.
Please follow the chart below to carry out battery performance
maintenance.
Figure 10-1. Flow chart of Battery Performance Maintenance
NOTE: Disconnect all probes when discharge battery.
NOTE: Discharge the battery to let the system automatically shut down.
Battery discharge
Battery charge > 3 hour
Yes
No
Battery power > 90%?
Battery charge > 3 hour
Battery power > 90%?
No
Battery is Ok
Yes
Yes
Discharge Replace battery
No
Yes
Battery is Ok
Replace battery
>45 minutes?
Replace battery
>45 minutes?
No