Site Preparations
2-10 Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
Probes environmental requirements
Operation and storage temperatures for probes
NOTE: Systems and electronic probes are designed for storage
temperatures of -5 to +50°C. When exposed to large
temperature variations, the product should be kept in room
temperature for 10 hours before use.
Table 2-7: Operation and storage temperatures for probes
Conditions Temperature
Operation: 10 to 40°C (50 to 104°F)
Storage: -5 to 50°C (23 to 122°F)
Temperature in degrees Celsius (°C) conversion to degrees F (°F):
(°F) = (°C * 9/5) + 32