System Setup
3-22 Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
Preparing for Setup
Verify customer order
Compare items received by the customer to that which is listed
on the delivery order. Report any items that are missing, back
ordered, or damaged.
Physical inspection
Verify that the system arrived intact (visual inspection).
If the system has been damaged, please refer to ‘Damage in
Transportation’ on page i-11 in the beginning of this manual.
System Voltage Settings
• Verify that the scanner is set to the correct voltage. The
Voltage settings for the Venue 50 Scanner is found on a
label located on the AC adapter.
• 220-240VAC(China); 100-120VAC(USA/Japan);
220-240VAC(Europe, Latin America).
NOTE: Check your local grid and confirm the voltage.
EMI protection
The Venue 50 has been designed to minimize the effects of
Electro-Magnetic Interference (EMI). Many of the covers,
shields, and screws are provided primarily to protect the system
from image artifacts caused by this interference. For this reason,
it is imperative that all covers and hardware are installed and
secured before the unit is put into operation.
See ‘EMI limitations’ on page 2-8 for more information about
EMI protection.
Connecting Docking Station/Cart to the wrong voltage
level will most likely destroy the system.