Vscan Extend Maintenance
6-8 Vscan ExtendUser Manual
5721203-100
Rev. D
Cleaning agents
Follow the instructions described below while using CaviWipes
and Sani-Cloth HB.
CaviWipes
1. Use one CaviWipes towelette to completely preclean the
surface of the device, probe and cable.
2. Ensure that no debris and bioburden are left behind.
3. Discard used towelette.
4. Use a second CaviWipes towelette to thoroughly wet the
surface in order to disinfect the precleaned surface.
Repeated use of the product may be required to ensure that
the surface remains visibly wet for 3 minutes at room
temperature (20 degrees).
Sani-Cloth HB
1. Use one Sani-Cloth HB towelette to completely preclean the
surface of the device, probe and cable
2. Ensure that no debris and bioburden are left behind.
3. Discard used towelette.
4. Use a second Sani-Cloth HB towelette to thoroughly wet the
surface in order to disinfect the precleaned surface.
Repeated use of the product may be required to ensure that
the surface remains visibly wet for 10 minutes at room
temperature (20 degrees).