EasyManua.ls Logo

GE Vscan Extend - Disinfecting the Device

GE Vscan Extend
251 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cleaning and disinfection
Vscan ExtendUser Manual 6-7
5721203-100 Rev. D
Disinfecting the device
After cleaning, the device may be wiped with a tissue sprayed
with a recommended germicide.
Recommended germicides
In order to provide users with options in choosing a germicide,
GE routinely reviews new medical germicides for compatibility
with the Vscan Extend and its probe. Although a necessary step
in protecting patients and employees from disease transmission,
liquid chemical germicides must also be selected to minimize
potential damage to the probe.
NOTE: Refer to the cleaning agent labeling for preparation and use
instructions.
NOTE: Follow the manufacturer's instructions for storage, use and
disposal of the disinfection solution.
NOTE: No special accessories are required for disinfection.
Table 6-1: Germicides
PI-Spray Pharmaceutical Innovations
PI-Spray II Pharmaceutical Innovations
CaviWipes Metrex
Cleanisept wipes Dr. Schumacher GmbH
Septiwipes Dr. Schumacher GmbH
Sani-Cloth HB PDI
WARNING
Use only compatible germicides. In addition, refer to the local /
national regulations.