Vscan Extend Maintenance
6-10 Vscan Extend – User Manual
5721203-100
Rev. D
Disinfecting the probe (continued)
Figure 6-2. Probe immersion level
WARNING
Do not immerse the probe in any liquid beyond the level
specified (see Figure 6-3).
The probe should not be exposed to the germicide longer than
specified in order to achieve the desired effect.
Rinse the part of the probe which is in contact with the
germicide according to the germicide manufacturer’s
instructions.
Wipe dry with a soft towel.
1. Maximum saline level
2. Minimum saline level
a. Linear array transducer
b. Phased array transducer
c. G3S probe
WARNING
Use additional precautions (e.g. gloves and gown) when
decontaminating an infected probe.
The probe should not be exposed to the germicide longer than
specified in order to achieve the desired effect.