Troubleshooting
Vscan ExtendUser Manual 6-15
5721203-100 Rev. D
Vscan Extend display is flashing
while scanning.
Automatic reduction of the frame
rate due to increase of the
operating temperature after
extended scanning.
Restart the Vscan Extend to
restore the normal frame rate.
NOTE: To maintain the operating
temperature of the Vscan Extend
at an optimal functional level, it is
recommended to hold the Vscan
Extend so there is good contact
between the device and the hand.
This will also ensure a longer
scanning time with maximum
frame rate
Vscan Extend is not entering the
scan screen and stays on the GE
start-up screen.
Battery may not have sufficient
charge.
Charge the Vscan Extend battery
for at least 60 minutes.
Table 6-2: Troubleshooting (Continued)