8-12 Vscan Extend – User Manual
5721203-100
 Rev. D
Local Archive - Security capabilities 
The Vscan Extend is provided with an internal archive, for 
storing images and patient data locally on the system. The local 
archives file repository and patient database do not support file 
sharing or remote connection. These can only be accessed 
locally. 
The patient database is protected with the authentication for 
access requirements and no possibility of remote access.
DICOM connections – Security capabilities
The DICOM connection works as defined by DICOM guidelines. 
The application accepts connection only to/from DICOM entities 
with IP-address, AE Title and port number matching the 
configured parameters in the Vscan Extend Ultrasound system.
The communication sessions are on demand and are always 
initiated locally from the system.
The Vscan Extend Ultrasound system’s internal firewall has 
exemptions for ports used by the defined DICOM dataflows in 
the system. Defining a new DICOM dataflow, or changing an 
existing dataflow, will cause the internal firewall configuration to 
automatically change. This ensures that only ports configured 
for a dataflow have an exemption in the internal firewall.
Windows share
Windows share access can be secured by defining a dedicated 
user on the server side. The user credentials for the network 
share user must be entered in the configuration UI on the Vscan 
Extend Ultrasound system.
NOTE:  There is no network file share on the Vscan Extend Ultrasound 
system.
Network infrastructure
The infrastructure of the network where the Vscan Extend is 
connected must be configured to allow traffic as described in 
Inbound firewall configuration and Outbound firewall 
configuration sections. All other traffic to and from the Vscan 
Extend can be blocked in the network infrastructure to prevent 
unintended access.