Safety
2-16 Vscan ExtendUser Manual
5721203-100
Rev. D
Electromagnetic immunity (continued)
Table 2-3: Electromagnetic immunity (Part 2)
Guidance and manufacturer’s declaration – electromagnetic immunity – for all medical
electrical equipment and medical electrical systems that not life-supporting
The Vscan Extend is intended for use in the electromagnetic environment below. The
customer or the user of the Vscan Extend should assure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Complianc
e level
Electromagnetic
environment – guidance
c
Conducted RF
EN 61000-4-6
Radiated RF
EN 61000-4-3
3Vrms
150 kHz to 80 MHz
3 V/m
80 Mhz to 2,5 GHz
3V
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the ultrasound unit, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
where p is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.

Table of Contents