EasyManua.ls Logo

GE XMO2

GE XMO2
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
September 2007
6-36 General Programming
CALIBRATE RECORDER
Menu
The Calibrate Recorder? menu is used to scale the XMO2’s 4-20 mA
analog output to any range within the maximum calibrated range of
your unit. Refer to the menu map in Figure C-6 on page C-6, and
proceed as follows:
Note: The 4-20 mA analog output range can also be programmed in
the Basic Menu, via the Set Low Input Value (page 5-10) and
Set High Input Value (page 5-11) options.
At the next prompt, enter the oxygen percentage that corresponds to a
4 mA recorder analog output.
At the next prompt, enter the oxygen percentage that corresponds to a
20 mA recorder analog output.
This completes the prompts for the Calibrate Recorder? option, and
the XMO2 returns to the main General menu.
GENERAL MENU
Press [Y] or [Enter] to calibrate
the recorder, or press
[N] to move
to the next General Menu option.
Calibrate Recorder?
mA Output 4 mA Value
To accept the current value, press
[Y] or [Enter]. To change the
current value, use the numeric
keys to enter a new value and
press
[Y] or [Enter] twice.
%O2 [x.xx]:
mA Output 20 mA Value
To accept the current value, press
[Y] or [Enter]. To change the
current value, use the numeric
keys to enter a new value and
press
[Y] or [Enter] twice.
%O2 [xxx.xx]:

Table of Contents

Related product manuals