Specifications and Factory Data 8-9
September 2007
Advanced Menu Defaults The default settings for the options in the Advanced Menu are:
Setup - Terminal
• Select Gas Type - [%O2]
• Display Background Gas - [No]
• Select Pressure Type - [MMHG]
• Set Easy Menu Entry - [Enable]
Calibrate System - System Mode
• Response Type - [Damped]
• Pressure Comp - [Off]
• Background Comp - [Off]
• Calibration Range - [0.00% - 100.00%?]
Calibrate System - O2 Input
• Enter AutoZero Value - [-400.0]
• AutoZero Bridge - [no default setting]
Calibrate System - DriftCal Handler
• Max % Total Drift - [25% F.S.]
• Max % Drift per DriftCal - [10% F.S.]
Calibrate System - Calibrate Recorder
• Enter mA Output Zero - [4.00]
• Enter mA Output Span - [20.00]
Set Error Handling - Terminal
• DriftCal Offset Error Response - [Enable]
• DriftCal Zero Error Response - [Enable]
• DriftCal Span Error Response - [Enable]
• Gas Input Underrange Response - [Enable]
• Gas Input Overrange Response - [Enable]
• Bkgd Input Underrange Response - [Enable]